6016Z
Release time:2021-02-05 19:01:44viewed:times
Introduction to the 6016 z particle counter
The dust particle counter is a application in the cleanliness of clean environment monitoring, testing instruments in the field of micro particle count test, etc. Counter based on the theory of light scattering theory, through the photoelectric sensing technology for detecting of dust particle size and particle number counting. The equipment is widely used in pharmaceutical research and development, the electronic circuit production, such as medical operating room clean. The instrument has good stability, high precision, small volume, light weight, easy installation, etc.
The main technical parameters:
粒径范围: 0.3~10.0μm
粒径通道: 0.5μm,5.0μm(默认)
采样流量: 28.3L/min(1CFM) ±5%
通讯协议: RS485 Modbus-RTU协议(9600 N1) 8
粒径分布误差: ≤±30%
浓度示值误差: ≤±30%FS
自净时间: ≤10min
重复误差: ±10%F.S
激光源: JinKouBanDaoTiJiGuangErJiGuan
真空要求: WaiZhiDuLiBeng
工作电压: DC24V
工作电流: 1A
工作条件: 温度10~40°C 湿度10%~90%无凝结
无凝结、 -10~50°C 存储条件: temperature humidity 90%
校准规范: JJF1190-2008
外壳: stainless steel
尺寸: 17.5 cm(长) cm(宽) x 13.6 x 6.8 cm(高)
Communications specification
parameters
Baud rate: 9600 bits: 8 check digit: no stop bit: 1
deal
Modbus RTU: -
03 function code to read the register
06 function code to write a single register
Note: the device address is 1 ~ 250, the broadcast address 254 (only when the 485 bus only one instrument can communicate using radio address test)
寄存器表(Modbus协议寄存器表可根据客户需求定制)
The Modbus address read/write data type description
0 to read, write an Integer instrument start-stop (1: start; 0: stop)
1 read the Integer count time (in seconds)
2 read low 16-bit Integer 0.3 um observation data
3 read high 16 bit Integer 0.3 um observation data
4 read low 16-bit Integer 0.5 um observation data
5 read high 16 bit Integer 0.5 um observation data
6 read low 16-bit Integer 1.0 um observation data
7 read high 16 bit Integer 1.0 um observation data
8 read Integer 3.0 um observation data