Hello, welcome to Suzhou gold instrument manufacturing co., LTD

FYJ - 2 planktonic bacteria sampler

Release time:2021-02-03 15:09:16viewed:times
FYJ - 2 planktonic bacteria sampler

A, product introduction
FYJ - 2 air suction flow rate is 100 litres per minute. Each sampling cycle, most can be collected 9999 liters of air. This performance can make the sample in accord with monitoring under strict sterile room and clean room. Impact velocity (speed) on the surface of the microorganism in the air crash AGAR around 10 m/s, is equivalent to the level 5 of Andersen sampler.
Note: FYJ - 2 reference standards: ISO 14698-1/2, GMP, GB16293-2010
Sampling speed to 100 l + 2.5%, using petri dish for 90-100 - mm diameter.
Basic principles:
FYJ - 2 is a high performance of the instrument. It use the principle of Andersen air sampler pumps air through the porous cover (465 * 0.6 mm), microbial adhesion, impact of airflow on the surface of standard AGAR medium in a petri dish. After complete the sampling period, put the dishes in appropriate conditions, and then determine the total number of colonies.
2. The main advantages
1) fit 90-100 - mm diameter standard AGAR, without having to use expensive "matching" special materials
2) built-in flow calibration, temperature and humidity calibration function
3) the external bluetooth printer, you can print the data information (time, date, temperature and humidity, sampling volume)
4) take export data software, generate EXCEL spreadsheets, convenient save data computer
5) adjustable sampling head support clips to fit a variety of plate size difference (90-100 - mm)
6) can be connected 100/240 v, 50 to 60 KHZ working power supply, or only use the batteries work
7) sample volume can be set arbitrary value between 1-9999 - l
8) USB connection way to ensure that the result records accurate
9) automatic sampling precision calibration, avoid artificial operation error



3. Technical parameters
1. The sampling head for anodic alumina, suitable for most of the disinfection method
2. Sampling flow rate: 100 l/min plus or minus 2.5%

The sampling time delay: 0-255 - s (adjustable)
Sample volume: 0-9999 - l (adjustable)
Impact speed: 10.8 m/s (Anderson impact rank the fifth principle)
Battery life: more than 5 hours
Dimension: 120 * 120 * 175 mm (D * W * H) weight: 2.1 Kg
Execution standard: ISO 14698-1/2, GB/T 16293-2010, GMP

Suzhou gold instrument manufacturing co., LTDSweep WeChat yards pay attention to us

  • 24-hour hotline<15006201004

  • The mobile phone15006201004

Copyright © 2019-2021 Central air conditioning co. LTD All rights reserved 地址:Jiangsu suzhou industrial park, east room 216, building no. 32 A rich road ICP:苏ICP备2024096266号-1